These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Detection and measurement of steady-state evoked potentials in real-time using a lock-in amplifier. Technical note. Author: Schacham SE, Pratt H. Journal: J Neurosurg; 1985 Jun; 62(6):935-8. PubMed ID: 3998849. Abstract: A system for recording evoked potentials, in which a lock-in amplifier replaces the averaging computer, is described. With this system, the set-up is much the same as in the conventional apparatus, with the following exceptions: 1) the amplifier gain is set at only a few thousands; 2) analogue-to-digital conversion and computer averaging is replaced by a lock-in amplifier; and 3) the output of the system is an analogue signal representing the amplitude of the recorded steady-state evoked potentials and the phase of baseline crossings. This set-up is much less costly than the conventional apparatus, much easier to operate, and, with its real-time output, may provide an ideal technique for screening and monitoring purposes.[Abstract] [Full Text] [Related] [New Search]