These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Performance evaluation of xenon-133 inhalation rebreathing systems for regional blood flow measurements. Author: Dunn WL, Oswald WM, Wahner HW. Journal: J Nucl Med; 1985 Jul; 26(7):783-9. PubMed ID: 4009287. Abstract: A quality-control phantom is described which is capable of testing the performance of inhalation rebreathing systems for measurements of regional cerebral blood flow (rCBF). The phantom is designed so that data similar to those from a patient study are obtained. The standard rCBF data-processing algorithm is used for the calculation of phantom study results. Malfunctions were induced to simulate instrument malfunction, and their effects on detector and air curves were evaluated. The use of this phantom is recommended prior to a patient study if the rCBF instrument is used infrequently or after service for routine maintenance or malfunction.[Abstract] [Full Text] [Related] [New Search]