These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Preparation and use of serum-based material as control and calibrator in evaluating ion-selective electrodes for calcium.
    Author: Toffaletti J, Lee KM.
    Journal: Clin Chem; 1985 Aug; 31(8):1349-52. PubMed ID: 4017239.
    Abstract:
    We analyzed 93 sera and 81 whole-blood samples, plus several aqueous materials and one serum-based material as controls, with two analyzers for ionized calcium (Radiometer ICA 1 and AVL 980). On 16 days, comparisons of results for patients' samples were good, with nearly all between-instrument differences (delta Ca2+) of samples being within 0.04 mmol/L. On 12 days, comparisons were mediocre, with the delta Ca2+ of samples usually 0.04 to 0.10 mmol/L. The control material that most consistently indicated the direction and magnitude of the delta Ca2+ of patients' samples was the serum-based control. Adjustment of results for samples from patients, based on the delta Ca2+ of the serum control on the same day, substantially improved comparisons between instruments. Our findings suggest that the use of a serum-based calibrator provides the most valid comparison between different calcium-ion analyzers and reliably indicates when electrode replacement is needed.