These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Respiratory ordered phase encoding (ROPE): a method for reducing respiratory motion artefacts in MR imaging. Author: Bailes DR, Gilderdale DJ, Bydder GM, Collins AG, Firmin DN. Journal: J Comput Assist Tomogr; 1985; 9(4):835-8. PubMed ID: 4019854. Abstract: A method of reducing respiratory artefact when using the spin warp technique of magnetic resonance imaging is described. A respiratory signal is used to determine the order in which the rows of the data matrix are measured. The aim is to make the respiratory signal at the time of each phase encoding gradient a slowly varying function of the time integral of the gradient and hence of the row number. Data are collected during each cycle of the imaging sequence; thus the scanning time is not increased.[Abstract] [Full Text] [Related] [New Search]