These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Experience with outlet strut fracture of the Björk-Shiley convexoconcave mitral valve prosthesis. Author: Mok CK, Lee JW, Kong SM, Hui KK. Journal: Am Heart J; 1985 Oct; 110(4):814-8. PubMed ID: 4050654. Abstract: Within a period of 15 months, we encountered five patients with outlet strut fracture of the Björk-Shiley convexoconcave mitral valve prosthesis. All of the patients had acute failure of the left side of the heart and pulmonary edema. The diagnosis was made by review of the chest roentgenogram which showed features of pulmonary edema; both the outlet strut and the tilting disc were missing from the mitral prosthesis. The "missing" components might be found in the left side of the heart or in the course of the thoracic aorta. Emergency reoperation was carried out on three patients, and there were two survivors. It is essential to recognize this potentially fatal complication and undertake immediate surgical intervention. We performed the original valvular replacement operations in four of the five patients. The probability of outlet strut fracture occurring in our series of 237 patients with a Björk-Shiley convexoconcave mitral valve prosthesis is 1.69%, or 0.076 per 100 patient months.[Abstract] [Full Text] [Related] [New Search]