Title: Improved estimation of fructosamine, as a measure of glycated serum protein, with the Technicon RA-1000 analyzer. Author: San-Gil F, Schier GM, Moses RG, Gan IE. Journal: Clin Chem; 1985 Dec; 31(12):2005-6. PubMed ID: 4064290. Abstract: We present an improved fructosamine assay for use with the Technicon RA-1000 analyzer. Features of this assay include: specimen throughput of 100 per hour, within-batch CV of 1.5%, between-batch CV of 2.4%, standard curve linear up to 24.9 mmol of fructosamine per liter, and good correlation (r = 0.82) with hemoglobin A1.[Abstract] [Full Text] [Related] [New Search]