These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Improved estimation of fructosamine, as a measure of glycated serum protein, with the Technicon RA-1000 analyzer.
    Author: San-Gil F, Schier GM, Moses RG, Gan IE.
    Journal: Clin Chem; 1985 Dec; 31(12):2005-6. PubMed ID: 4064290.
    Abstract:
    We present an improved fructosamine assay for use with the Technicon RA-1000 analyzer. Features of this assay include: specimen throughput of 100 per hour, within-batch CV of 1.5%, between-batch CV of 2.4%, standard curve linear up to 24.9 mmol of fructosamine per liter, and good correlation (r = 0.82) with hemoglobin A1.
    [Abstract] [Full Text] [Related] [New Search]