These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: The variable reagent blank: protein determination as a model.
    Author: Bozimowski D, Artiss JD, Zak B.
    Journal: J Clin Chem Clin Biochem; 1985 Oct; 23(10):683-9. PubMed ID: 4067517.
    Abstract:
    Three total protein assays were analyzed to determine the extent of deviation encountered when a constant measured reagent blank is compared to a continuously decreasing true reagent blank. This blank effect owes its regressive nature to the consumption of the active reagent ingredient by the protein reactive species, variably and sometimes, with certain reactants, nonlinearly in the presence of increasing protein concentrations. However, the blank effect of interest here is apparent only when the reagent and the reagent-protein complex present overlapping spectra and therefore absorb at the wavelength of measurement. Thus it was found that while the biuret and the Coomassie brilliant blue assays produced pronounced differences in the variable true reagent blanks, the Folin-Ciocalteau reaction did not develop a deviation from the true blank since the reagent blank does not absorb to any extent at the assay wavelength. In this manner, the latter procedure could serve as a marker against which the former two blank reactions can be shown to display relatively excessive deviations.