These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Determination of erythromycin in serum and urine by high-performance liquid chromatography with ultraviolet detection.
    Author: Stubbs C, Haigh JM, Kanfer I.
    Journal: J Pharm Sci; 1985 Oct; 74(10):1126-8. PubMed ID: 4078714.
    Abstract:
    A high-performance liquid chromatographic analysis of erythromycin in human serum and urine with UV detection at 200 nm is presented. The method involves a solid-phase extraction procedure followed by a simple phase separation step and chromatography on a reversed-phase column. The method has sensitivity limits of 0.25 and 1.0 micrograms/mL in serum and urine, respectively, and is sufficiently sensitive to monitor concentrations of erythromycin in human serum and urine after the administration of a single 500-mg erythromycin stearate tablet.