These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Direct assays of lactate, pyruvate, beta-hydroxybutyrate, and acetoacetate with a centrifugal analyzer.
    Author: Hansen JL, Freier EF.
    Journal: Clin Chem; 1978 Mar; 24(3):475-9. PubMed ID: 415828.
    Abstract:
    Methods are described for direct assays of lactate, pyruvate, beta-hydroxybutyrate, and acetoacetate in plasma with the GEMSAEC centrifugal analyzer. The methods for lactate, beta-hydroxybutyrate, and acetoacetate are kinetic and ratiometric, eliminating the need for specimen-blank assays. The pyruvate method is an end-point assay, because endogenous lactate dehydrogenase interferes in a kinetic pyruvate assay. The methods are precise and accurate and 1-min of analysis time is adequate for each assay. Rapid assessment and monitoring of metabolic acidosis is possible with these methods, as is illustrated by examples.