These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Semiquantitative use of Biocept-G radioreceptorassay. Author: Hussa RO, Mattingly RF. Journal: Obstet Gynecol; 1979 Aug; 54(2):199-204. PubMed ID: 460754. Abstract: Serum samples (93 total) from 62 different patients were analyzed by quantitative radioimmunoassay (RIA) and by the Biocept-G radioreceptorassay (RRA) for detection of human chorionic gonadotropin (hCG). The ratio of sample counts to reference control counts (S/R) was determined for each RRA. For the 44 serum samples having S/R ratios between 0.295 and 1.20 (corresponding to hCG levels of 4600 and 89 mIU/ml, respectively), there was good correlation between the S/R ratio by RRA and the serum level of hCG by RIA (correlation coefficient = -0.92). When serum samples were analyzed in as many as 4 different Biocept-G kits, the same S/R ratios were obtained. The S/R ratio reliably predicted the trend in serum hCG titer in 7 patients during treatment for gestational trophoblastic disease. The usefulness of the S/R ratio in obtaining semiquantitative results with the Biocept-G RRA is discussed.[Abstract] [Full Text] [Related] [New Search]