These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Manipulative and inherent errors in anti-D quantitation using the AutoAnalyzer. Author: Gunson HH, Phillips PK, Stratton F. Journal: J Clin Pathol; 1972 Mar; 25(3):198-205. PubMed ID: 4622996. Abstract: The limits of accuracy for the quantitation of anti-D using the AutoAnalyzer have been reduced from +/- 26% with manually prepared dilutions to +/- 14% using an automatic pipette, as expressed by the 95% confidence limits. The error inherent in the AutoAnalyzer was estimated to contribute +/- 10% to the overall error. Problems associated with the reproducibility of this method for anti-D quantitation have been investigated, namely, the effect of the age of a given test cell, the use of different test cells, loss of sensitivity of the machine over a given day, and the reproducibility of results obtained at different positions on the standard graph.[Abstract] [Full Text] [Related] [New Search]