Title: Evaluation of the BMC glucose oxidase/peroxidase-4-aminophenazone-phenol procedure for glucose as adapted to the Technicon SMAC. Author: Purcell GV, Behenna DB, Walsh PR. Journal: Clin Chem; 1979 Oct; 25(10):1844-6. PubMed ID: 476940. Abstract: We evaluated the analytical performance of Trinder's glucose oxidase (EC 1.1.3.4)/peroxidase (EC 1.11.1.7) 4-aminophenazone-phenol method for the quantification of serum glucose as adapted to the Technicon SMAC. Our results correlated well with those by the routine SMAC glucose oxidase/peroxidase 3-methyl-2-benzothiazolinone hydrazone-N,N-dimethylaniline method (y = 1.02x - 49.4; r = 0.99) and the glucose oxidase oxygen-rate method (y = 0.99x + 14; r = 0.99) with the Beckman Glucose Analyzer. Sample-to-sample interaction was less than 1%. Ascorbic acid or uric acid in concentrations as high as 200 mg/L were without demonstrable effect on results for glucose. Intra- and inter-assay precisions (CV) were 1.6 and 2.3%, respectively. The upper limit of linearity was about 5 g/L. Adaptation of the Trinder method for glucose to the SMAC is simple and provides an analytically acceptable and economical alternative to the methods ordinarily used with the SMAC.[Abstract] [Full Text] [Related] [New Search]