These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Determination of manganese in whole blood and serum. Author: Pleban PA, Pearson KH. Journal: Clin Chem; 1979 Nov; 25(11):1915-8. PubMed ID: 498501. Abstract: We describe methods for determination of manganese in whole blood and serum with Zeeman-effect flameless atomic absorption spectroscopy. These analyses are performed on a twofold or fourfold dilution of the specimen in Triton X-100, 1 g/L. No predigestion or extraction procedures are required. The method of standard additions was used for quantitation. Within-run coefficients of variation for whole-blood manganese were 7.0 and 5.5% for 2.29 and 5.67 micrograms/L, respectively. For determination of serum manganese, coefficients of variation were 10.3 and 5.3% for 0.97 and 3.01 micrograms/L, respectively. Manganese detection limits for the assays were 3.0 pg. Whole-blood manganese concentrations, determined for 60 subjects, yielded a mean (+/- SD) of 9.03 (+/- 2.25) micrograms/L. Mean serum manganese concentration, determined for 20 subjects, was 1.82 (+/- 0.64) microgram/L. No correlation was found between blood manganese concentrations and age, sex, or smoking status.[Abstract] [Full Text] [Related] [New Search]