These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Is antiglycolysis required for routine glucose analysis?
    Author: Sazama K, Robertson EA, Chesler RA.
    Journal: Clin Chem; 1979 Dec; 25(12):2038-9. PubMed ID: 509703.
    Abstract:
    We obtained -68 pairs of simultaneously drawn serum and fluoride-oxalate plasma samples from patients and analyzed them by a continuous-flow (AutoAnalyzer II) glucose oxidase method. Glucose concentrations ranged from 370 to 3530 mg/L. Glucose concentrations for samples obtained in the fluoride-oxalate preservative averaged 42 +/- 35 mg/L (mean +/- SD) higher than serum. The magnitude of this difference was independent of glucose concentration. Linear-regression analysis of 270 pairs for which the time from collection to separation was recorded indicated that the difference between serum and plasma increased by 0.32 mg/L per minute of delay over a time span of 15 to 295 min. These differences are smaller than those described in standard textbooks. We conclude that, with the specimen-handling process used in our hospital, serum glucose determinations are clinically acceptable.
    [Abstract] [Full Text] [Related] [New Search]