These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: [Measurement of cardiac output by thermodilution with a diode as a temperature sensor].
    Author: Díaz Fernández A, Benítez D, Sánchez Tello G, Márquez LA.
    Journal: Prensa Med Mex; 1979; 44(3-4):45-52. PubMed ID: 550124.
    Abstract:
    An area integrator for the thermodilution curve in cardiac output measurement is described. A new temperature sensor is used, a diode with some advantages over the thermistor normally used. The main advantages are: easy calibration and replacement, and broad range of linearity. The cardiac output values obtained in dog with the integrator follow a linear relationship with those of the flowmeter. In simultaneous measurements the correlation is R = 0.96. Using a diode as temperature sensor a modification of the Steward Hamilton equation (used for thermistor) is necessary. With this new equation a monogram is performed to calculate the cardiac output from the area given by the numerical integrator.
    [Abstract] [Full Text] [Related] [New Search]