These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: The Humphrey Vision Analyzer tm: reliability and validity of refractive-error measures.
    Author: Kratz LD, Flom MC.
    Journal: Am J Optom Physiol Opt; 1977 Oct; 54(10):653-9. PubMed ID: 605920.
    Abstract:
    Four subjective measures of the refractive error of 42 eyes (21 patients) were obtained by 2 examiners, one using a Humphrey Vision Analyzer TM and the other using a conventional refractor. Each examiner made 2 separate measures of each subject's refractive error, one without and one with an added cylindrical lens chosen at random from a predetermined set of powers and axes. The order of using the instruments and the added lenses was mixed. Measures of the refractive errors differed with the 2 instruments by approximately the same amount on average as did duplicate measures with either instrument. Refractive-error measures obtained with the 2 instruments are about as valid as they are repeatable.
    [Abstract] [Full Text] [Related] [New Search]