These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Atrial systole and ineffective pacemaker stimuli. Author: Goicolea De Oro A, Ayala GS, Ayza MW, Meabe JA, Díez JR, Garrido RH, Ripoll AS, Alvarez JG. Journal: Pacing Clin Electrophysiol; 1981 Sep; 4(5):517-22. PubMed ID: 6169036. Abstract: An unusual case of pacemaker failure induced by atrial activity is presented. The pacemaker stimulus was ineffective for ventricular capture not only with low energy output (decreased pulse duration) of the generator, but also with pulse width two to three times the stimulation threshold. The two possible mechanisms of this phenomenon, one electrotonic and the other mechanical, are discussed. This report strongly supports the mechanical hypothesis.[Abstract] [Full Text] [Related] [New Search]