These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: An optimized collimator for single-photon computed tomography with a scintillation camera. Author: Kircos LT, Leonard PF, Keyes JW. Journal: J Nucl Med; 1978 Mar; 19(3):322-3. PubMed ID: 632913. Abstract: The problem of nonuniform response, which is inherent in scintillationcamera emission computed tomography, can be minimized by proper collimator design. A specially fabricated collimator, optimized for emission computed tomography, was designed. This collimator has a calculated full width at half maximum of 1 cm at 15 cm in front of the collimator. When this collimator is used in conjunction with averaged opposed views, excellent uniformity across the reconstruction plane is achievable.[Abstract] [Full Text] [Related] [New Search]