These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Evaluation of two methods of self blood glucose monitoring by trained insulin-dependent diabetic adolescents outside the hospital. Author: Schiffrin A, Desrosiers M, Belmonte M. Journal: Diabetes Care; 1983; 6(2):166-9. PubMed ID: 6343020. Abstract: We studied the accuracy of the Chemstrip bG and Glucometer systems in the self-monitoring of blood glucose by trained adolescents. The determinations were done at home with simultaneous collection of whole blood into capillary tubes (Sarstedt) which were later analyzed by a glucose-oxidase analyzer (Beckman Instruments). In both cases, there was an excellent correlation between laboratory concentrations and Chemstrip bG (r = 0.96, P less than 0.001) and Glucometer (r = 0.96, P less than 0.001). Comparisons made at 8 mo remained with the same degree of accuracy. There was a trend toward greater deviation with higher plasma glucose values. Well-trained patients can achieve sufficient accuracy to permit the use of either of the methods tested with similar results.[Abstract] [Full Text] [Related] [New Search]