These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Resolution in electron microscope autoradiography. III. Iodine-125, the effect of heavy metal staining, and a reassessment of critical parameters.
    Author: Salpeter MM, Fertuck HC, Salpeter EE.
    Journal: J Cell Biol; 1977 Jan; 72(1):161-73. PubMed ID: 63463.
    Abstract:
    Resolution for 125I-labeled specimens under electron microscope (EM) autoradiographic conditions was assessed experimentally. With this isotope the size of the silver halide crystal was the most important resolution-limiting factor. Heavy metal staining such as is routinely used in preparing animal tissues for EM autoradiography produced an improvement in resolution of approximately 15-20%. For a 500-1,000-A biological tissue section fixed with OsO4 and stained with uranyl acetate, we obtained resolution (half distance, HD) values of approximately 800 +/- 120 A using Ilford L4 emulsion and 500 +/- 70 A using a Kodak NTE-type emulsion. General aspects of resolution-limiting factors and comparison with 3H and 14C values are discussed.
    [Abstract] [Full Text] [Related] [New Search]