These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Semi-automated continuous-flow enzyme immunoassay for antiepileptic drugs in serum. Author: Hayashi S, Kurooka S, Arisue K, Kohda K, Hayashi C. Journal: Clin Chem; 1983 Oct; 29(10):1790-2. PubMed ID: 6352086. Abstract: We have developed a semi-automated method for measuring five kinds of antiepileptic drugs in serum by successfully adapting commercial competitive-binding enzyme immunoassay kits (MARKIT; Dainippon) for use with a continuous-flow analyzer (Technicon AutoAnalyzer II equipped with a dialyzer). The free enzyme-labeled drug is automatically separated by a microfilter from the competitive immunoreaction mixture between labeled and unlabeled drug for anti-drug immunoglobulin coupled to bacterial cell walls. The concentrations of the antiepileptic drugs in serum samples can be determined by automated measurement of enzyme activity of the enzyme-labeled drugs. Results of the semi-automated method correlated well with those obtained by manual enzyme immunoassay, gas-liquid chromatography, and "high-pressure" liquid chromatography. The correlation coefficients were all greater than 0.95, showing the practicality of this method for therapeutic monitoring of antiepileptic drugs.[Abstract] [Full Text] [Related] [New Search]