Search MEDLINE/PubMed
Title: Effect of volume and rate of contrast medium injection on intravenous digital subtraction angiographic contrast medium curves. Author: Burbank FH, Brody WR, Bradley BR. Journal: J Am Coll Cardiol; 1984 Aug; 4(2):308-15. PubMed ID: 6376594. Abstract: The image quality of temporal (mask mode) intravenous digital subtraction angiography is directly dependent on the shape of arterial time-concentration curves produced by the intravenous injection of contrast medium. Curves that are narrow and tall minimize motion artifact (misregistration) and maximize contrast enhancement (pre- and postcontrast differences). To determine the effects of rate and volume of injection of contrast medium on intravenous digital subtraction angiographic curves, ioxaglate (Hexabrix), a monoacidic ionic dimer, was injected into large mongrel dogs. Quantitative measurements of opacification were made over time in the femoral arteries using a modified General Electric CT/T scanner. Peak opacification was directly proportional to the volume of contrast medium injected. Curve width was not affected by increasing volume of injection. At rates below a critical point, slower injection rates produced progressively shorter and wider arterial time-concentration curves. Above that critical point, increasing the rate of injection did not affect either curve width or curve peak.[Abstract] [Full Text] [Related] [New Search]