These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Analysis of digoxin concentrations in serum by fluorescence polarization immunoassay: an evaluation.
    Author: Ferreri LF, Raisys VA, Opheim KE.
    Journal: J Anal Toxicol; 1984; 8(3):138-40. PubMed ID: 6376949.
    Abstract:
    The Abbott TDx fluorescence polarization immunoassay was evaluated for the determination of serum digoxin concentrations. Within-assay precision was less than 4% coefficient of variation (CV) for concentrations ranging from 0.64 to 3.75 ng/mL. Between-assay precision was 14.5% CV at 0.75 ng/mL, 5.7% CV at 1.50 ng/mL, and 4.9% CV at 3.48 ng/mL. Sensitivity to 0.2 ng/mL digoxin was confirmed. Correlation of 86 patient specimens assayed by radioimmunoassay (RIA) with the TDx showed the following: correlation coefficient r = 0.94, slope = 0.93, intercept = 0.11, and Sy/x = 0.19. Recovery from serum at concentrations of 0.97 ng/mL and 4.50 ng/mL averaged 98%. No significant interference from lipemia, icteria, or hemolysis was observed. Spironolactone showed no cross-reactivity with the antibody, while digitoxin exhibited significant cross-reactivity. Compared to the RIA procedure, the TDx assay was more rapid, reliable and, in this clinical situation, more cost effective.
    [Abstract] [Full Text] [Related] [New Search]