  • Title: Collaborative evaluation of the Abbott Avantage system for identification of frequently isolated nonfermentative or oxidase-positive gram-negative bacilli.
    Author: Jorgensen JH, Dyke JW, Helgeson NG, Cooper BH, Redding JS, Crawford SA, Andruszewski MT, Prowant SA.
    Journal: J Clin Microbiol; 1984 Nov; 20(5):899-904. PubMed ID: 6392323.
    Abstract:
    The capability of the Abbott Avantage system to identify 10 species of commonly isolated glucose nonfermentative or oxidase-positive gram-negative bacilli in a 5-h test period was evaluated in a collaborative study. The Avantage nonenteric data base uses 20 biochemical test reactions performed in an expanded Abbott bacterial identification cartridge plus the results of a manual oxidase test. The species included in the Avantage data base are Acinetobacter anitratus, Acinetobacter Iwoffi, Aeromonas hydrophila, Flavobacterium meningosepticum-IIb group, Pseudomonas aeruginosa, Pseudomonas cepacia, Pseudomonas fluorescens-putida group, Pseudomonas maltophilia, Pasteurella multocida, and Plesiomonas shigelloides. The collaborative study included the testing of 200 coded challenge strains in all three laboratories and the subsequent testing of an additional group of 100 to 200 clinical isolates recovered independently by each laboratory. Reference identifications for all isolates were determined by conventional biochemical test reactions. The overall accuracy of identification of the coded challenge strains for the three laboratories was 97%, whereas 95% of 437 clinical isolates and selected stock cultures of clinical derivation were identified correctly.
    [Abstract] [Full Text] [Related] [New Search]