These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Contaminant electrons in the build-up region of a 4 MV photon beam. Author: Rustgi SN, Gromadzki ZC, Ling CC, Yorke ED. Journal: Phys Med Biol; 1983 Jun; 28(6):659-65. PubMed ID: 6410419. Abstract: The per cent ionisation curves for the 4 MV photon beam from a Clinac 4 exhibit pronounced enhancement with field size in the build-up region. The source and nature of this increase is investigated by analysing the radiation components of the 4 MV photon beam with a 0.23 T magnetic field. The build-up ionisation curves were measured at 86 cm SSD with a parallel-plate ionisation chamber embedded in a polystyrene phantom. The surface ionisation increases from 18% to 32% as the field size is increased from 4.3 cm X 4.3 cm to 9.7 cm X 32.3 cm. With the electromagnet placed beside the treatment head, the field size dependence reduces to 16% to 24%. This and similar results at 113.2 cm SSD, clearly demonstrate that electrons from the Clinac 4 treatment head are responsible for approximately half the observed increase in the surface ionisation with field size.[Abstract] [Full Text] [Related] [New Search]