These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Duplex sonography in the evaluation of carotid artery disease. Author: Dreisbach JN, Seibert CE, Smazal SF, Stavros AT, Daigle RJ. Journal: AJNR Am J Neuroradiol; 1983; 4(3):678-80. PubMed ID: 6410830. Abstract: High-resolution duplex sonography was compared with biplane magnified carotid angiography in a prospective evaluation of 161 carotid arteries in 86 patients. The duplex scanner combined real-time B-mode imaging (7.5 MHz) with simultaneous range-gated pulsed Doppler frequency analysis (3 MHz). The degree of stenosis was usually determined by the true and residual lumen of the carotid artery at the atherosclerotic plaque on the transverse image. The Doppler frequency signals were automatically converted to velocity data by a minicomputer. The accuracy of the duplex system in detecting and assessing stenoses graded in 20% increments is demonstrated according to specificity, sensitivity, and positive and negative predictive value by both cumulative and subgroup analyses. A 93.7% sensitivity was shown for minimal (0-20%) stenosis and 100% sensitivity for severe (greater than 60%) stenosis. An apparent limitation of duplex sonography is the differentiation of a high-grade stenosis from occlusion (sensitivity, 82.6%; positive predictive value, 90.4%).[Abstract] [Full Text] [Related] [New Search]