These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Fluorometric measurement of serum magnesium with a centrifugal analyzer.
    Author: Pesce MA, Bodourian SH, Hills LP.
    Journal: Clin Chim Acta; 1984 Jan 31; 136(2-3):137-44. PubMed ID: 6420091.
    Abstract:
    We developed a rapid fluorometric assay for measurement of serum magnesium using the ligand, 8-hydroxy-quinoline-5-sulfonate and adapted the procedure to the Multistat Fluorescence Light Scattering Centrifugal Analyzer. The standard curve extends from 0.26 to 4.11 mmol/l. There was no interference from calcium or inorganic phosphorus at concentrations of 4.95 and 5.0 mmol/l, lead, iron, zinc or copper at twice the normal levels found in serum, bilirubin at concentrations of 10 mg/dl, or lipemic samples with triglyceride concentrations of 2400 mg/dl. Citrate and EDTA lowered magnesium concentrations in serum. Analytical recovery of magnesium added to four serum specimens averaged 97%. Between-run and within-run precision of the assay gave CVs which ranged from 2.9 to 7.6%. Magnesium concentrations, measured by our fluorometric procedure, were compared with those obtained by atomic absorption and colorimetric procedures. Correlation coefficients of 0.91 and 0.88 were obtained.
    [Abstract] [Full Text] [Related] [New Search]