These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Endotracheal cuff pressure and tracheal mucosal blood flow: endoscopic study of effects of four large volume cuffs.
    Author: Seegobin RD, van Hasselt GL.
    Journal: Br Med J (Clin Res Ed); 1984 Mar 31; 288(6422):965-8. PubMed ID: 6423162.
    Abstract:
    Large volume, low pressure endotracheal tube cuffs are claimed to have less deleterious effect on tracheal mucosa than high pressure, low volume cuffs. Low pressure cuffs, however, may easily be overinflated to yield pressures that will exceed capillary perfusion pressure. Various large volume cuffed endotracheal tubes were studied, including Portex Profile, Searle Sensiv, Mallinkrodt Hi-Lo, and Lanz. Tracheal mucosal blood flow in 40 patients undergoing surgery was assessed using an endoscopic photographic technique while varying the cuff inflation pressure. It was found that these cuffs when overpressurised impaired mucosal blood flow. This impairment of tracheal mucosal blood flow is an important factor in tracheal morbidity associated with intubation. Hence it is recommended that a cuff inflation pressure of 30 cm H2O (22 mm Hg) should not be exceeded.
    [Abstract] [Full Text] [Related] [New Search]