These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Calibration of some Wright peak flow meters. Author: Fisher J, Shaw A. Journal: Br J Anaesth; 1980 Apr; 52(4):461-4. PubMed ID: 6445741. Abstract: The performance of Wright peak flow meters is described using an absolute and precise calibration method. This showed errors as great as 25% with the peak flow meter and 40% with the mini peak flow meter. At large flow rates errors were less. The expected variation between individual peak flow meters was about 12%. The expected variation between individual mini peak flow meters was between 16% and 22%, depending on flow rate.[Abstract] [Full Text] [Related] [New Search]