These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Calibration of some instruments for measuring peak expiratory flow.
    Author: Shaw A, Fisher J.
    Journal: J Med Eng Technol; 1980 Nov; 4(6):291-4. PubMed ID: 6453234.
    Abstract:
    The performance of instruments for measuring peak expiratory flow have been investigated. This evaluation was carried out with two designs of spirometer, two types of pneumotachograph and three versions of the Wright peak flow meter. The appropriate flow profiles were generated by means of a special type of flow calibrator. The errors produced by the instruments varied from zero to about 26%.
    [Abstract] [Full Text] [Related] [New Search]