These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Evaluation of immunoassays for cannabinoids in urine. Author: Irving J, Leeb B, Foltz RL, Cook CE, Bursey JT, Willette RE. Journal: J Anal Toxicol; 1984; 8(4):192-6. PubMed ID: 6471820. Abstract: A comparison was made of several cannabinoid urine assays. Two hundred randomly selected urine specimens were initially screened by two enzyme immunoassays (EMIT-st and EMIT-d.a.u.) and a radioimmunoassay (Abuscreen RIA). Selected specimens found positive by any of these methods were further analyzed by gas-liquid chromatography with flame ionization detection (GLC/FID), gas chromatography/mass spectrometry (GC/MS), and an experimental RIA from Research Triangle Institute (RTI RIA). The GLC/FID method gave confirmations in 69 to 92% of the samples, depending on the method used and the cut-off employed. GC/MS confirmed 98% of the EMIT and RIA positives using a low cut-off (20 ng/mL). All RIA positives at 100 ng/mL were confirmed by GC/MS. There was complete agreement between the RTI RIA and the EMIT assays, but not with the Abuscreen RIA at the 100 ng/mL cut-off. The study illustrates that care must be exercised in establishing assay cut-offs and the designation of false positive results.[Abstract] [Full Text] [Related] [New Search]