  • Title: The use of cadmium telluride detectors for the qualitative analysis of diagnostic x-ray spectra.
    Author: Di Castro E, Pani R, Pellegrini R, Bacci C.
    Journal: Phys Med Biol; 1984 Sep; 29(9):1117-31. PubMed ID: 6483976.
    Abstract:
    A method is introduced for the evaluation of x-ray spectra from x-ray machines operating in the range 50-100 kVp using a cadmium telluride (CdTe) detector with low detection efficiency. The pulse height distribution obtained with this kind of detector does not represent the true photon spectra owing to the presence of K-escape, Compton scattering, etc.; these effects were evaluated using a Monte Carlo method. A stripping procedure is described for implementation on a Univac 1100/82 computer. The validity of our method was finally tested by comparison with experimental results obtained with a Ge detector and with data from the literature; the results are in good agreement with published data.