These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: High pressure liquid chromatographic determination of strychnine, using a reverse phase solvent system.
    Author: Crouch MD, Short CR.
    Journal: J Assoc Off Anal Chem; 1978 May; 61(3):612-5. PubMed ID: 649554.
    Abstract:
    A simple, rapid extraction and subsequent determination for strychnine, using high pressure liquid chromatography (HPLC) with a reverse phase solvent system, is described. Stomach contents or grain bait containing strychnine were made alkaline with sodium hydroxide and extracted with chloroform. Extract filtrates were injected directly into a liquid chromatograph without further preparation except for dilution, if necessary. Peaks were resolved within 3.5 min and peak heights were used for quantitation. HPLC of strychnine was carried out on a 30 cm X 4 mm id stainless steel column packed with micronBondapak C18. The solvent program was 0.005M phosphate buffer-methanol (60+40) at a flow of 1.5 ml/min. Recovery from spiked stomach contents was 93.9 +/- 3.5%. The detection capability for strynchnine, using the 254 nm ultraviolet detector, was 5 ng. The strychnine peak was collected and subjected to thin layer chromatography with strychnine standards for confirmation.
    [Abstract] [Full Text] [Related] [New Search]