These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Potential pitfalls in the use of Glucoscan and Glucoscan II meters for self-monitoring of blood glucose.
    Author: Laus VG, Dietz MA, Levy RP.
    Journal: Diabetes Care; 1984; 7(6):590-4. PubMed ID: 6510183.
    Abstract:
    We discovered that skilled nurses only casually trained in the use of a fingertip blood glucose reflectance meter (Glucoscan, Lifescan, Mountainview, California) had a 36% incidence of unacceptable results (greater than 15% from reference). A controlled study was undertaken and showed that with Glucoscan I (GI) 4 of 27 readings were unacceptable and with Glucoscan II (GII) 3 of 27 readings were unacceptable, a statistically nonsignificant difference. Minor deviations from the manufacturer's recommended technique had a significant effect on the results with GI. In contrast, GII was much less sensitive to variations in recommended technique. GI underestimated the reference glucose concentration by 11.7%, and GII overestimated by 6.5%, a statistically significant difference. We conclude that the health professional must be aware of interdevice and intradevice variability in self-monitoring of blood glucose (SMBG). Patients need careful training in the method of SMBG. The results of any single value should be interpreted with caution.
    [Abstract] [Full Text] [Related] [New Search]