These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: [Blood glucose determination at the bedside with test strips and reflection photometer: comparison of 2 instruments with a standard method].
    Author: Cretti M, Gloor U, Albrecht R, Bürgi H.
    Journal: Schweiz Med Wochenschr; 1984 Oct 27; 114(43):1492-4. PubMed ID: 6515348.
    Abstract:
    The accuracy of blood glucose measurements was evaluated with two different test strip-reflection photometer systems (Reflolux and Reflocheck). Two experienced technicians took 92 measurements in capillary blood at the bedside and the results were compared with a reference value obtained from the same capillary blood sample by a standard automated glucose oxidase method. Both test strips agreed well with the reference value, with high correlation coefficients and with total errors of less than 25%. It remains to be established whether the two methods are as reliable when performed by patients themselves.
    [Abstract] [Full Text] [Related] [New Search]