These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Formulas for preparing phantom muscle tissue at various radiofrequencies. Author: Chou CK, Chen GW, Guy AW, Luk KH. Journal: Bioelectromagnetics; 1984; 5(4):435-41. PubMed ID: 6517962. Abstract: The dielectric properties of various test samples of phantom tissue were measured using an automated and temperature-controlled slotted line. The ingredients for phantom materials were determined for simulating high-water content tissue at 13.56, 27.12, 40.68, 70, 100, 200, 300, 433, 750, 915, and 2,450 MHz. The ingredients consisted of water, TX-150 (a gelling agent), sodium chloride, and polyethylene powder (200-2,450 MHz) or aluminum powder (13.56-100 MHz). The dielectric constant and conductivity of these materials at different temperatures (15, 22, 30 degrees C) were characterized.[Abstract] [Full Text] [Related] [New Search]