These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Determination of salbutamol in human serum by reversed-phase high-performance liquid chromatography with amperometric detection. Author: Tan YK, Soldin SJ. Journal: J Chromatogr; 1984 Nov 28; 311(2):311-7. PubMed ID: 6520178. Abstract: A simple and sensitive method for the quantitative determination of salbutamol in human serum using reversed-phase ion-pair high-performance liquid chromatography with electrochemical detection is described. The method involves the combined use of Sep-Pak cartridges and ion-pair extraction for sample clean-up, and subsequent separation of salbutamol and the internal standard from interfering compounds on a reversed-phase column. An amperometric detector incorporating a glassy carbon electrode was employed for detection. The inter-assay coefficients of variation at plasma concentrations of 2.0, 6.0 and 20.0 ng/ml were 7.3%, 7.2% and 8.5%, respectively (n = 20). The minimum detection limit was 400 pg/ml from a 0.5-ml sample of serum. The method can be readily utilised for clinical pharmacokinetic studies.[Abstract] [Full Text] [Related] [New Search]