These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Prolonged dialyzer reuse.
    Author: Kaye M, Gagnon R, Mulhearn B, Spergel D.
    Journal: Trans Am Soc Artif Intern Organs; 1984; 30():491-3. PubMed ID: 6533928.
    Abstract:
    Clearance of urea, creatinine, phosphate and B12 in dialyzers reused up to 30 times were unchanged from the initial values. Residual formalin values were within acceptable levels and in several thousand dialyses no untoward reactions of any sort attributable to reuse were encountered. Neutropenia was constantly present with dialyzers prepared by automated reuse.
    [Abstract] [Full Text] [Related] [New Search]