These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Computer-controlled anesthesia delivery system.
    Author: Hayes JK, Westenskow DR, East TD, Jordan WS.
    Journal: Med Instrum; 1984; 18(4):224-31. PubMed ID: 6548544.
    Abstract:
    A prototype Computer-Controlled Anesthesia Delivery System (CCADS) has been developed. The CCADS uses proportional, integral, and derivative (PID) software algorithms to control the delivery of oxygen, anesthetic agent,and nitrous-oxide (N2O) via the computer feedback control of precision mass-flow controller devices. The CCADS was evaluated in a group of seven dogs during 4 hours of closed-circuit enflurane N2O anesthesia. Throughout the 4-hour monitoring period, the CCADS maintained the animals' inspired oxygen concentration (FIO2), end-tidal enflurane concentration, and breathing circuit volume to within +/- 0.20, +/- 0.10 volume %, and +/- 30 ml/min (+/- 1 SD), respectively, of the desired values. The interactive software allowed selective graphic display of the 10-breath averaged data parameters, or modification of FIO2, end-tidal enflurane concentration, and breathing circuit volume, whenever animal re-evaluation during an anesthesia procedure was necessary.
    [Abstract] [Full Text] [Related] [New Search]