These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: A rapid micro-scale method for the measurement of Haemoglobin A1(a+b+c). Author: Welch SG, Boucher BJ. Journal: Diabetologia; 1978 Mar; 14(3):209-11. PubMed ID: 658636. Abstract: A rapid method is described for the measurement of total glycosylated haemoglobins (HbA1(a+b+c). The procedure utilizes 0.05 ml of blood and takes forty minutes to complete manually. Eighty blood samples can be analysed without automation by one person in a day. Each analysis uses less than 2 mg of potassium cyanide, resulting in a method that is both safe and rapid for routine hospital laboratories. The inter-assay coefficient of variation was 4% and that for intra-assay measurements 3%, over the range 5-20%, HbA1(a+b+c). The method confirmed that the level of HbA1(a+b+c) is elevated in imperfectly controlled diabetics. Amongst patients with blood glucose levels of less than 10 mmol/l the mean level of HbA1(a+b+c) was found to be 8.5%; samples from 14 known diabetics gave a mean value of 10.9%, whereas 17 known non-diabetic samples gave a mean value of 8.3%. In the group of samples from 27 diabetic individuals with blood glucose levels above 10 mmol/l the mean level of HbA1(a+b+c) was found to be 13.5%.[Abstract] [Full Text] [Related] [New Search]