These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: The effect of temperature on plasma ionized calcium measured in whole blood in vitro. Author: Heining MP, Band DM, Linton RA. Journal: Scand J Clin Lab Invest; 1983 Dec; 43(8):709-14. PubMed ID: 6665519. Abstract: Plasma calcium ion concentration (cCa2+) in samples from 50 healthy volunteers was measured at four temperatures (21, 26, 31 and 37 degrees C) using a calcium electrode based on a neutral carrier ligand. A small negative correlation was found between temperature and cCa2+, the coefficient being -0.0017 mmol . l-1 . degrees C-1 (P less than 0.001). The significance of this in clinical, physiological and instrumentation fields is discussed. The effect on measured cCa2+ of overnight storage at 4 degrees C was slight, but did increase the variance of the results. It is suggested that the most accurate results of cCa2+ will be obtained from fresh samples measured at body temperature.[Abstract] [Full Text] [Related] [New Search]