These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: A neodymium-YAG endolaser.
    Author: Peyman GA, Conway MD, Ganti S, Viherkoski E.
    Journal: Ophthalmic Surg; 1983 Apr; 14(4):309-13. PubMed ID: 6683392.
    Abstract:
    We used a CW-YAG endolaser to photocoagulate the fundus of the rabbit and monkey. We varied the power settings, exposure durations, and probe-retina distances to establish the parameters for threshold and bursting (choroidal hemorrhage) lesions. We found that the amount of energy needed to create a retinal lesion increased proportionally with the distance between the probe and the retina. To produce a choroidal hemorrhage at a given distance required at least four times the power required for a threshold coagulative lesion at the same distance. We also found that we could achieve a strong choroidal reaction without hemorrhage by using the coagulative threshold parameters of power and distance with an increased exposure time. Photocoagulation of the retina using the CW-YAG depended on the presence of melanin.