These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Heat and mass transfer of a thermal indicator in pulsatile flow through the cardio-pulmonary system. II. Identification of cardiac output. Author: Cassot F, Saadjian A, McKay C. Journal: Med Prog Technol; 1978 May 23; 5(4):215-222. PubMed ID: 672825. Abstract: Hamilton's celebrated formula for cardiac output measurement is simple but its validity is dependent on several methodologic requirements which are not generally fulfilled, particularly in thermal dilution. A quite different method, based on a physico-mathematical model of the indicator dispersion in the circulation, is proposed. It allows direct derivation of cardiac output once the model's parameters have been identified. Combined deconvolution and least squares procedures are used with truncated data for this identification. Numerical tests and application to clinical observations are presented. Both limitations and possibilities of further developments in estimation of pulsatile flow conditions from TD technique are discussed.[Abstract] [Full Text] [Related] [New Search]