These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Discordant inter-kit results in the radioimmunoassay for choriogonadotropin in serum. Author: Rzasa PJ, Caride VJ, Prokop EK. Journal: Clin Chem; 1984 Jul; 30(7):1240-3. PubMed ID: 6733907. Abstract: Discordant results for concentrations of choriogonadotropin (HCG) in sera from five of 162 women are reported, as determined with four commercial radioassay kits: Becton-Dickinson, Clinetics, Corning, and Hybritech. Results ranged from 11.2 to 36.7 int. units/L, when assayed with the Becton-Dickinson kit, but HCG was not detectable when assayed with the other kits. None of the patients was pregnant, and none had evidence of persistent trophoblastic or nontrophoblastic disease. Concentrations of lutropin, follitropin, thyrotropin, triglycerides, cholesterol, protein, and albumin were normal. Assay of diluted sera did not parallel the standard curve, and nonspecific binding in the patients' samples was negligible. Some kits for HCG may be influenced by inherent methodological factors or interfering serum constituents, which can lead to spurious results.[Abstract] [Full Text] [Related] [New Search]