These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Evaluation of the MS-2 urine screening method for detection of bacteriuria. Author: McCarthy LR, Gavan TL, Robson J, Corlett C. Journal: J Clin Microbiol; 1982 Aug; 16(2):250-2. PubMed ID: 6749888. Abstract: A collaborative evaluation of the MS-2 urine screening system indicated that the automated method could successfully identify at least 84% of urine specimens that contained greater than or equal to 10(5) colony-forming units per ml. Most specimens not detected as positive by the system contained primarily diphtheroids and yeasts. Excluding those specimens which contained diphtheroids and yeasts, the MS-2 system correctly identified as positive 94.8% of specimens that contained greater than or equal to 10(5) colony-forming units per ml and as negative 96.1% of specimens that contained less than 10(4) colony-forming units per ml. The automated system is an attractive method for the screening of urine specimens for significant bacteriuria.[Abstract] [Full Text] [Related] [New Search]