These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Respiratory compensation during spontaneous ventilation with the Bain circuit.
    Author: Byrick RJ.
    Journal: Can Anaesth Soc J; 1980 Mar; 27(2):96-105. PubMed ID: 6767540.
    Abstract:
    The volume of carbon dioxide rebreathed by spontaneously breathing patients under halothane anaesthesia at various fresh gas flow rates (FGF) with the Bain modification of the Mapleson "D" breathing circuit is measured. The effect of rebreathing on a heterogeneous patient population is shown to be unpredictable hypercapnia in those patients who cannot respond adequately to this carbon dioxide challenge. All adults rebreathe significant volumes of carbon dioxide at a FGF rate of 100 ml . kg-1 . min-1. This carbon dioxide load is a potential risk to every patient and this hypercapnia is preventable by using high FGF rates. Rebreathing occurs because the inspired carbon dioxide load is unpredictable in a given patient and the patient's response is uncontrolled. Patients respond to this carbon dioxide challenge by increasing inspiratory flow rate (Vt/Ti), which results in increased rebreathing of carbon dioxide from the expiratory limb of the circuit. To prevent potentially dangerous rebreathing of carbon dioxide in all patients the fresh gas flow rate must be much higher than presently recommended.
    [Abstract] [Full Text] [Related] [New Search]