These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Effect of ventilation and diffusion nonuniformity on DLCO (exhaled) in a lung model.
    Author: Cotton DJ, Graham BL.
    Journal: J Appl Physiol Respir Environ Exerc Physiol; 1980 Apr; 48(4):648-56. PubMed ID: 6769882.
    Abstract:
    Recent studies have shown that diffusing capacities measured at multiple intervals during a single exhalation [DLCO(exhaled)] remained constant with lung volume in normal subjects, but decreased with decreasing lung volume in patients who may have had diffusion nonuniformity. We have examined the theoretical basis of these results by determining what factors affected DLCO(exhaled) in a computerized lung model in which diffusion in each compartment remained constant with lung volume. DLCO(exhaled) decreased with decreasing lung volume when a small lung region lacked diffusion. However, the change in DLCO(exhaled) with lung volume was also affected by nonuniform ventilation and these effects could not be eliminated by correcting the carbon monoxide decay and the simultaneous decay of helium. DLCO(exhaled) values were also influenced by the exhaled flow rate in the presence of nonuniform ventilation and/or nonuniform diffusion. However, prolonging the period of breath holding prior to exhalation reduced DLCO(exhaled) values at all lung volumes when non-uniform diffusion was simulated, but did not affect DLCO(exhaled) when only nonuniform ventilation was simulated.
    [Abstract] [Full Text] [Related] [New Search]