These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Noninvasive transcutaneous carbon dioxide monitoring.
    Author: Hebrank DR, Mentelos RA.
    Journal: Med Instrum; 1981; 15(3):203-6. PubMed ID: 6792474.
    Abstract:
    A new method for continuously and noninvasively monitoring carbon dioxide pressure in the skin tissue using an unheated transcutaneous carbon dioxide sensor is presented. The design, calibration, and principle of operation of the sensor are discussed. The transcutaneous PCO2 sensor was tested in vitro for response time, drift, linearity, and interfering anesthetic gases, all of which showed that the sensor is accurate, reliable, and immune to anesthetic gases. Clinical results using the unheated transcutaneous PCO2 sensor with neonatal patients are presented, verifying the clinical usefulness of the device for noninvasively monitoring trends in arterial PCO2.
    [Abstract] [Full Text] [Related] [New Search]