Search MEDLINE/PubMed
Title: Thin layer chromatographic determination of aflatoxin in corn dust. Author: Shotwell OL, Burg WR, Diller T. Journal: J Assoc Off Anal Chem; 1981 Sep; 64(5):1060-3. PubMed ID: 6793549. Abstract: Methods adopted by the AOAC and the American Association of Cereal Chemists for determining aflatoxin in corn were modified, and techniques were developed for application to samples of less than 1 to 10 g instead of the specified 50 g samples. Analysis included chloroform extraction of dust samples or dust collected from glass fiber filters, purification of extracts on a silica gel column of appropriate size, and measurement of aflatoxin by either 1- or 2-dimensional thin layer chromatography (TLC). The solvent for 1-dimensional TLC was chloroform-acetone-water (91 + 9 + 1). Solvents for 2-dimensional TLC were, first direction, ether-methanol-water (95 + 4 + 1, lined tank) and second direction, chloroform-acetone-water (91 + 9 + 1, unlined tank), or first direction, chloroform-acetone-water (91 + 9 + 1, unlined tank) and second direction, toluene-ethyl acetate-formic acid (60 + 30 + 10, unlined tank). When samples weighed less than or equal to 0.1 g, the entire concentrated extract was applied to the TLC plate. About 0.5-1.0 ng aflatoxin B1 could be detected on the plate, making the limit of detection about 9 ng/g for 0.1 g samples.[Abstract] [Full Text] [Related] [New Search]