These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Determination of urinary normetanephrine and metanephrine by radial-compression liquid chromatography and electrochemical detection.
    Author: Orsulak PJ, Kizuka P, Grab E, Schildkraut JJ.
    Journal: Clin Chem; 1983 Feb; 29(2):305-9. PubMed ID: 6821934.
    Abstract:
    A procedure has been developed for determining the O-methylated catecholamine metabolites, normetanephrine and metanephrine, in urine by use of radial-compression liquid chromatography followed by electrochemical detection. Normetanephrine and metanephrine are isolated from hydrolyzed urine by ion-exchange on small, commercially available, disposable columns and preconcentrated by solvent extraction. They are then separated by reversed-phase ion-pair chromatography, with use of a radial compression cartridge and radial compression module, and quantified with 3-methoxy-4-hydroxybenzylamine as internal standard. Normetanephrine, metanephrine, and the internal standard are separated from interfering peaks in about 15 min. The method is applicable to the relatively low amounts of normetanephrine (100-600 micrograms/24 h) and metanephrine (50-400 micrograms/24 h) found in normal subjects and patients with depressive disorders or hypertension. Within-day CVs ranged from 1.1 to 2.2% for normetanephrine and 1.2 to 6.9% for metanephrine; the corresponding between-day CVs were 4.9 and 5.7% over these ranges.
    [Abstract] [Full Text] [Related] [New Search]